Description

The Alpha 300RA is a combines TrueSurface Raman imaging and AFM imaging in one instrument.

Raman

  • Laser wavelength: 532 nm up to ~ 55 mW.
  • Resolution: 1 cm-1 or ~ 3 cm-1 depending on the grating.
  • High sensitivity EM CCD.
  • 360nm lateral spatial resolution with 100x objective (0.9N.A.).
  • excellent confocality and depth resolution (<1 um).
  • 50×50 mm stage.
  • large Z range.
  • Reflected light.
  • image stitching.
  • Truesurface to compensate for sample topography while imaging the sample.

AFM

  • resolution X/Y ~3 nm and Z ~0.5 nm.
  • 80×80 um maximum area.
  • 5 um Z range.
  • AC (tapping) and contact mode.
  • force-distance curves.
  • TERS capable with appropriate tips.

Contact

Emmanuel Soignard
Operations Director
[email protected]
480.965.7242

Techniques

  • Optical spectroscopy
  • Surface imaging

Rates

Service or materialASU rateNonprofit or academic rateNotes
Equipment use$31.00/hr$65.00/hr
Equipment use with staff assistance$91.00/hr$161.00/hr